| Product Characteristics |
| Number of Inputs and Outputs | Number of Analog Outputs: 2 |
| Process Connection | Diameter Ø 8 mm |
| System | Gold-plated contacts |
| Media | Water; glycol solutions; air; oils |
| Medium temperature [° F] | -4 - 212 |
| Pressure Rating [bar] | 50 |
| Electrical Data |
| Operating Voltage [V] | 18 - 30 DC |
| Current Consumption [mA] | < 100 |
| Protection Class | III |
| Reverse Polarity Protection | Yes |
| Power-on Delay Time [s] | 10 |
| Inputs / Outputs |
| Number of Inputs and Outputs | Number of Analog Outputs: 2 |
| Outputs |
| Total Number of Outputs | 2 |
| Output Signal | Analog signal |
| Number of Analog Outputs | 2 |
| Analog Current Output [mA] | 4 - 20; (scalable) |
| Maximum Load [Ω ] | 350 |
| Short-Circuit Protection | Yes |
| Type of Short-Circuit Protection | Yes (non-latching) |
| Overload Protection | Yes |
| Measuring/Setting Range |
| Probe Length L [mm] | 200 |
| Operating Mode | Relative; absolutely liquid; absolutely gaseous; (absolute: reference measurement recommended; Factory Setting: relative) |
| Liquids |
| Resolution [ft/s] | 0.05 |
| Analog start point ASP [ft/s] | 0 - 7.95 |
| Analog end point AEP [ft/s] | 1.9 - 9.85 |
| Gases |
| Resolution [ft/s] | 2 |
| Analog start point ASP [ft/s] | 0 - 264 |
| Analog end point AEP [ft/s] | 64 - 328 |
| Temperature Monitoring |
| Measuring range [° F] | -4 - 212 |
| Resolution [° F] | 0.5 |
| Analog start point [° F] | -4 - 169 |
| Analog end point [° F] | 39 - 212 |
| In steps of [° F] | 0.5 |
| Liquid Media - Absolute Operating Mode |
| Setting Range [ft/s] | 0 - 9.85 |
| Greatest sensitivity [ft/s] | 0.15 - 9.85 |
| Liquid media - Relative Operating Mode |
| Setting Range [ft/s] | 0 - 19.5 |
| Greatest sensitivity [ft/s] | 0.15 - 9.85 |
| Gases - Operating Mode " absolute" |
| Setting Range [ft/s] | 0 - 328 |
| Greatest sensitivity [ft/s] | 6 - 328 |
| Gases - Operating Mode " relative" |
| Setting Range [ft/s] | 0 - 656 |
| Greatest sensitivity [ft/s] | 6 - 328 |
| Accuracy / Deviations |
| Temperature Drift [cm/s x 1/K] | 0,01 fps x 1/K (<68 ° F; >158 ° F) |
| Maximum Temperature Gradient of Medium [K/min] | 100 |
| Absolute Operating Mode |
| Repeatability | 0,05 m/s; (water; Flow velocity: 0,05 - 3 m/s) |
| Relative Operating Mode |
| Accuracy | ± (7 % MW + 2 % MEW); (for relative mode in the range of maximum sensitivity under the following conditions: ; water: 68 - 158 ° F; inlet length: 5 ft; DN25 (DIN 2448); mounting position according to instructions; Accuracy can differ for other media and mounting positions.) |
| Repeatability | 0,05 m/s; (water; Flow velocity: 0,05 - 3 m/s) |
| Temperature Monitoring |
| Temperature Drift | ± 0,003 K/° F |
| Accuracy [K] | ± 0,3 / ± 1; (water; Flow velocity: 1 - 9,85 fps / air; Flow velocity: >32,8 fps) |
| Reaction Times |
| Response Time [s] | 0.5; (T09; water; glycol: 0,8 s; air: 7 s; oil: 1,8 s; each T09) |
| Temperature Monitoring |
| Dynamic Response T05 / T09 [s] | 1,5 (T09); (water; Flow velocity: 1 - 9,85 fps) |
| Software / Programming |
| Parameter Setting | Medium selection; Damping; Teach function; display can be rotated and switched off; standard unit of measurement; process value color |
| Operating Conditions |
| Ambient Temperature [° F] | -40 - 176 |
| Storage Temperature [° F] | -40 - 212 |
| Protection | |
| MTTF [Years] | 179 |
| Mechanical Data |
| Weight [g] | 310.5 |
| Material | Stainless steel (1.4404 / 316L); PBT-GF20; PBT-GF30 |
| Materials (Wetted Parts) | Stainless steel (1.4404 / 316L) |
| Process Connection | Diameter Ø 8 mm |
| Electrical Connection |
| Connection | Connector: 1 x M12; coding: A; Contacts: gold-plated |