PCE Instruments, PCE-TDS 100HSH, Ultrasonic Flow Meter in 3/4" - 28" Tubes
PCE-TDS 100HSH Includes Only:
- 1 x PCE-TDS 100H Ultrasonic flow meter
- 2 x Sensors TDS-M1
- 2 x Sensors TDS-S1
- 2 x 5 m / 16.4 ft transducer connection cables
- 4 x Velcro-strap clamps
- 3 x AAA Ni-H rechargeable batteries
- 1 x Battery charger
- 1 x Container of coupling gel
- 1 x Measuring tape
- 1 x Carrying case
- 1 x User manual
Features:
- Bi-directional flow measurement
- Quick and adjustable response time
- Eliminates interfering frequencies
- Includes data logger functionality
- Stores up to 2000 measurement values
- Auto power-off battery-saving function (can be disabled)
- Saves time- and date-stamped measurement data to internal memory
- Rechargeable, battery-operated flow measuring device
PCE-TDS 100HSH is a portable handheld clamp-on ultrasonic flow meter kit used for non-invasive, unobstructed, and highly accurate measurements of the flow velocity of liquids in metal, plastic, and rubber pipes and tubes with a diameter of 20 - 720 mm / approx. 3/4" - 28". Ideal for use in oil and gas, water and wastewater, chemical, food and beverage, pharmaceutical, metals and mining, pulp and paper, power and heating, ventilation, air conditioning, and refrigeration (HVACR) industries, this ultrasonic flow meter kit features user-friendly velcro-strap clamps that allow for quick and easy repositioning of the electroacoustic transducers.
Measurable liquids include acetate, acetone, alcohol, ammonia, aniline, benzene, butyrate, chloroform, ethanol, ethyl alcohol, ethyl ether, ethylene glycol, freon R 12, petrol, glycerin, glycol, isobutanol, isobutane, isopentane, kerosene, linseed oil, methanol, methyl alcohol, engine oil, diesel oil, olive oil, peanut oil, paraffin oil, pentane, petroleum, 1-propanol, coolant, lubricating oil, silicone oil, transformer oil, trichloroethylene, 1, 1, 1 - trichloroethane, turpentine, distilled water, and seawater.
Calculation of flow velocity according to the transit-time principle follows the equation:
v = measured velocityT1 = run time of the ultrasonic signal in the flow directionT2 = run time of the ultrasonic signal against the flow direction L = length of the ultrasonic waveα = ultrasonic signal angle to the flow
The transit-time principle requires pipes to be full and have no bubbles and no particles.
Note: To transfer data to a computer, SOFT-PCE-TDS software is required. This software is sold separately - see accessories for details.
Each PCE-TDS series meter is assembled by PCE Instruments in Germany and factory-calibrated (without any documentation). The reference display of the in-house test stand used by PCE for calibration has a valid DAkkS calibration certificate. This ensures traceability to the Physikalisch-Technische Bundesanstalt (PTB) German national standard. Please note that the meter's measured values depend on the pipe geometry, material, and coating; the medium type, temperature, and speed; and the sensor type and measuring method.