Product Characteristics |
Number of Inputs and Outputs | - Number of Digital Outputs: 1
- Number of Analog Outputs: 1
|
Probe Length L [mm] | 100 - 2000 |
Process Connection | Threaded connection 3/4" NPT external thread |
System | gold-plated contacts |
Media | Liquids |
Dielectric Constant of the Medium | >5 |
Recommended Media | Water; water-based media |
Process temperature [°C] | -25 - 80; (90<1 h) |
Pressure Rating [bar] | 16 |
Vacuum Resistance [mbar] | -1000 |
Electrical Data |
Operating Voltage [V] | 18 - 30 DC |
Current Consumption [mA] | <25 |
Protection Class | III |
Reverse Polarity Protection | Yes |
Power-on Delay Time [s] | <3 |
Measuring Principle | guided wave radar |
Inputs / Outputs |
Number of Inputs and Outputs | - Number of Digital Outputs: 1
- Number of Analog Outputs: 1
|
Outputs |
Total Number of Outputs | 2 |
Output Signal | Switching signal; analog signal; IO-Link |
Electrical Design | PNP/NPN |
Number of Digital Outputs | 1 |
Output Function | Normally open / closed; (configurable) |
Maximum Voltage Drop Switching Output DC [V] | 2.5 |
Permanent Current Rating of Switching Output DC [mA] | 200 |
Number of Analog Outputs | 1 |
Analog Current Output [mA] | 4 - 20, invertible |
Maximum Load [Ω] | 500 |
Analog Voltage Output [V] | 0 - 10, invertible |
Minimum Load Resistance [Ω] | 2000 |
Short-Circuit Protection | Yes |
Type of Short-Circuit Protection | Yes (non-latching) |
Overload Protection | Yes |
Measuring/Setting Range |
Probe Length L [mm] | 100 - 2000 |
Active Range A [mm] | L-40 |
Inactive Range I1 / I2 [mm] | 30 / 10 |
Sampling rate [Hz] | 4 |
Setting Range |
Set Point SP [mm] | 15 - L-30 |
Reset Point rP [mm] | 10 - L-35 |
In Steps of [mm] | 5 |
Hysteresis [mm] | >5 |
Accuracy / Deviations |
Repeatability [mm] | 5 |
Measuring Error [mm] | ± 7 |
Offset Error [mm] | 5 |
Resolution [mm] | 1 |
Zero Signal (Voltage) [V] | 0 |
Zero Signal (Current) [mA] | 4.0 |
Full Signal (Voltage) [V] | 10 |
Full Signal (Current) [mA] | 20 |
Temperature Drift per 10 K | ± 0.2 % |
Interfaces |
Communication Interface | IO-Link |
Transmission Type | COM2 (38,4 kBaud) |
IO-Link Revision | 1.1 |
SDCI Standard | IEC 61131-9 |
Profiles | Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis |
SIO Mode | Yes |
Required Master Port Class | A |
Process Data Analog | 3 |
Process Data Binary | 1 |
Minimum Process Cycle Time [ms] | 3.2 |
Operating Conditions |
Ambient Temperature [°C] | -25 - 60 |
Storage Temperature [°C] | -40 - 85 |
Protection | IP 68; IP 69K; (7 days / 1 m water depth / 0.1 bar: IP 68) |
MTTF [Years] | 241 |
Mechanical Data |
Weight [g] | 448.2 |
Material | - Stainless steel (1.4301 / 304)
- Stainless steel (1.4404 / 316L)
- FKM
- PEI
|
Materials (Wetted Parts) | - Stainless steel (1.4305 / 303)
- Probe connection: stainless steel (1.4435 / 316L)
- PTFE
- FKM
|
Process Connection | Threaded connection 3/4" NPT external thread |
Notes | For high process temperatures: The temperature at the Process Connection is decisive. The actual medium temperature may be higher. |
Electrical Connection - plug |
Connection | Connector: 1 x M12; coding: A; Contacts: gold-plated |