Subject to the use of all the optional accessories |
Resolution | 0.1 V |
50 Hz to 400 Hz +/-(% of Rdg + Digits) | +/-(2% + 3) |
Input impedance | 3 MΩ (nominal), < 100 pF |
Common mode rejection ratio (1 kΩ unbalanced) | > 60 dB at DC, 50 or 60 Hz |
Overload protection | 600 V rms or DC |
Range/Resolution | 20.00 Ω |
200.0 Ω | 0.1 Ω |
2000 Ω | 1.0 Ω |
20.00 kΩ | 0.01 kΩ |
Accuracy | +/-(1.5% + 3) |
Overload protection | 2 V rms or DC |
Open circuit test voltage | > 4.0 V, < 8 V |
Short circuit current | > 200.0 mA |
Measurement range | 0.01 MΩ to 2000 MΩ |
Test voltage | 500 V, 1000 V |
Test voltage accuracy | + 20%, - 0% |
Short circuit current | 1 mA nominal |
Auto discharge | Discharge time < 0.5 second for C = 1 μF or less |
Live circuit indicator | Inhibit test if terminal voltage > 30 V prior to initialization of test |
Maximum capacitive load | Operable with up to 1 μF load |
500 V | +/-(1.5% + 5) |
1000 V | +/-(1.5% + 5) to 2000 MΩ, +/-(10% + 3) above 2000 MΩ |
Maximum voltage applied to any terminal: | 600 V AC rms or DC |
Storage temperature | -40°C to 60°C (-40°F to 140°F) |
Operating temperature | -20°C to 55°C (-4°F to 131°F) |
Temperature coefficient | 0.05 x (specified accuracy) per °C for temperatures < 18°C or > 28°C (< 64°F or > 82°F) |
Vibration | Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument |
Shock | 1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor) |
Electromagnetic compatibility | In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-1:1997) |
Safety | Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement category IV 600 V (CAT IV) |
Certifications | CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition |
Batteries | Four AA batteries (NEDA 15A or IEC LR6) |
Resistance measurements: | Tester can perform at least 2500 earth bond resistance measurements with fresh alkaline batteries at room temperature. These are standard tests of 1 Ω with a duty cycle of 5 seconds on and 25 seconds off. |
Size (H x W x L) | 5.0 x 10.0 x 20.3 cm (1.97 x 3.94 x 8.00 in) |
Weight | 550 g (1.2 lb) |
IP rating | IP40 |
Non-operating (storage) | 12,000 m |
Over-range capability | 110% of range |
Included accessories | TL224 Test Leads, TP74 Test Probes, clips PN 1958654 (red) and PN 1958646 (black), holster and remote probe |